MK 11 S2 (635.011.2)

MK 11 S2Voltage gage head for direct voltage measuring up to ±2kV for the EFM 120. With circlip and BNC plug.

The MK11 S2 is mounted on the electro field meter's modulator system. As the teflonized measuring plate is located in a distance of 1cm, the following measuring ranges are obtained:

field intensity: voltage range:
±2kV/m ±20V
±20kV/m ±200V
±200kV/m ±2kV


These voltage meters have excellent properties:

Eingangskapazität: 5pF
Eingangswiderstand: >1015